Areas of Application
The device is to be used for carrying
out real-time biokinetic, immunosensing,
and biosensing measurements; studying
nanoobjects like nanoparticles or nanotubes;
studying adsorption, corrosion, and
electrochemical reactions; measuring refractive
index of organic and inorganic fi lms; detecting
gas and liquid composition; realizing
chemosensor applications; and controlling
motor oil quality and wear of meeting parts
Advantages
The device has no counterparts in Ukraine.
It requires low doses of substances to be tested
(in ml), is compact and cheaper than foreign
analogs
IPR Protection
IPR3
Specifi cation
Number of channels
Refractive index measurement
range
Refractive index sensitivity
Maximum time resolution
of kinetics measurements, s
Tracing measurement mode
Slope measurement mode
Angle-of-incidence precision,
angular s
GaAs semiconductor laser
as light source
Overall dimensions
of the measurement unit, mm
Weight, kg
2
1.0-1.45
0.00005
2
0.2
5
650 nm, 2-3 mW
215×130×100
2.5
Stage of Development.
Suggestions for Commercialization
IRL3, TRL4
Manufacture, delivery, warranty service,
and staff training, upon request